One option is to use heat-resistant tape. A roll of heat-resistant tape is stored in the red tool box.
An alternative is to use the green Mylar™ tape stored in the Parylene Coater box on a shelf in the main area of the cleanroom.
Scissors can be found in the red tool box.
Press RESET to acknowledge the error.
Press SEAL.
Press VENT and allow the chamber to fully vent before opening the door.
Close the chamber door again and press CYCLE.
The instrument restarts the pump down cycle and errors do not typically re-occur.
In the event that the above sequence does not fix the problem and another “PUMPDOWN TIMEOUT” error is generated:
Press RESET to acknowledge the error.
Press SEAL.
Press STOP and allow the turbo pump to decelerate.
Once the display shows STAND-BY, press on START and wait for the turbo pump to accelerate. Once the turbo pump is ready, the CYCLE button will be available.
Press CYCLE. The instrument will restart the pump down cycle.
Before using the basket for evaporation, it must be outgassed. Put the basket in place between the two electrodes (make sure that each end of the basket sits between two washers); tighten firmly but take care not to brake/snap the basket. Do not load any evaporant (metal) in the basket; do not load any samples. Proceed as if starting a run; when ready, slowly heat up the filament and maintain a current of at least 3A for 30min.
Use cleanroom wipe and IPA to wipe any area that shows peeling/delamination. Wipe the red seal on the inside of the door if metal residues are still visible. Make sure that you only use the IPA bottles located in the cleanroom; do not use the IPA bottle from the gowning room (70% dilution) to avoid introducing water in the system.
Use the vacuum cleaner again as more debris might have fallen during wipe out.
A FltrS error indicates that the filter in the front of the unit needs to be cleaned. The error occurs after 1000 hrs of use. It is manually reset when the filter is cleaned and re-installed.
The chiller will continue to function when the error message is displayed.
Please use contact information on this website to send a message to the instrument supervisor. The supervisor will clean and reset the chiller.